Gas Flow Calibration

Use the Gas Flow Constant Calibration option to determine a constant used by the analyzer to ensure accurate gas flows through each Mass Flow Controller (MFC). The Mass Flow Controllers were calibrated before shipping and calibration is typically not necessary.

  1. Go to Unit [n] > Enable Manual Control. Ensure a checkmark displays to the left of the menu item. If the analyzer schematic does not display, go to Unit [n] > Show Instrument Schematic. Flow the gas to be calibrated through MFC to be used for the calibration.
  2. Connect a flow meter to the exhaust and measure the actual flow rate.
  3. In the Gas drop-down box, select the gas to be flowed. In the Actual flow field, enter the actual flow rate to compute the mass flow constant.
  4. Click Accept to change the mass flow constant in the Adsorptive Properties for the gas to be the computed value.
 
   Micromeritics Learning Center    Contact Us

Copyright © 2018. Micromeritics Instrument Corporation. All rights reserved.