Transfer a Degassed Physical Adsorption Sample to an Analysis Port

The sample tube must be removed from the degas port, weighed, and then installed onto the analysis port for analysis.

  If the sample tube will not be mounted on the analysis port immediately, leave it on the degas port. If it is necessary to remove the sample tube and a Seal Frit was not used, insert a rubber stopper into the sample tube.

  1. Allow the sample tube to cool.
  Do not touch the sample tube or the heating mantle until they have reached room temperature. Touching the sample tube, heating mantle, or heating mantle clip before they have cooled could result in burns.
  1. Carefully remove the heating mantle clip and the heating mantle from the sample tube.
  2. Hold the sample tube, loosen the port connector nut, and remove the sample tube from the degas port.
  3. Remove the connector nut, ferrule, and O-ring from sample tube stem.
  4. Weigh the sample tube set. Use the Sample Data Worksheet to determine the sample mass.
  5. Slide an isothermal jacket down over the sample tube stem until it touches the sample tube bulb.
  6. Place the connector nut, ferrule, and O-ring onto the sample tube stem.
  7. Remove the stopper and immediately attach the sample tube to the analysis port, pushing it fully up into the port. Secure it in place by screwing the connector nut onto the analysis port connector; hand tighten the connector nut. If a Seal Frit was used, it does not have to be removed.
  8. Place the dewar cover over the sample tube stem just above the isothermal jacket.
 
   Micromeritics Learning Center    Contact Us

Copyright © 2014 - 2018. Micromeritics Instrument Corporation. All rights reserved.