Select a Sample Tube
A sample tube set consists of:
- Sample tube
- Stopper or Seal Frit
- Filler rod
Standard sample tubes for the
A rubber stopper may be used with all size sample tubes; however, seal frits are recommended for 1.27 cm (1/2 in.) OD sample tubes.
Filler rods help to ensure accuracy in samples with lower total surface areas by reducing the free space volume. It is generally a good practice to use filler rods for samples having less than 100 square meters of total surface area. Filler rods are unnecessary for samples with total surface areas greater than 100 square meters.
Filler rods can interfere with thermal transpiration correction and, therefore, should not be used when performing micropore analyses. |
The weight of the empty sample tube should be determined after it has been cleaned, degassed, and filled with backfill gas. The sample tube should be allowed to cool to room temperature before backfilling. After the sample tube has cooled, remove it from the degas port and weigh it.
If a Seal Frit is not used, insert a stopper immediately after removing the sample from the degas port. |
The mass of the isothermal jacket may vary slightly and cannot be considered as constant; therefore, do not weigh it with the sample tube set.
Copyright © 2014 - 2018. Micromeritics Instrument Corporation. All rights reserved.